In vitro electrochemical corrosion and biotribological behavior of flame-sprayed single and bilayer coatings on SLM-fabricated CoCrWMo alloy


BAHÇEPINAR A. İ., AYDIN İ.

SURFACES AND INTERFACES, cilt.101, 2026 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 101
  • Basım Tarihi: 2026
  • Doi Numarası: 10.1016/j.surfin.2026.111106
  • Dergi Adı: SURFACES AND INTERFACES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, INSPEC
  • Manisa Celal Bayar Üniversitesi Adresli: Evet

Özet

This study comparatively investigates the electrochemical and biotribological behaviors of single- and bilayer coating systems developed to enhance the biomedical surface performance of CoCrWMo implant alloys fabricated via Selective Laser Melting (SLM). To suppress potential metallic ion release and promote bone-implant interface integration, single-layer Titanium (Ti), Hydroxyapatite (HA), and dual-layer Ti/HA coatings were synthesized using the cost-effective and commercially scalable powder flame spray technique. Microstructural and phase characterizations revealed that reactive TiN and Ti3O ceramic phases formed during Ti deposition, whereas the Ti/HA bilayer structure exhibited a multi-phase mixture of HA, beta-TCP, and highly soluble TTCP, yielding a chemically favorable Ca/P ratio of 1.71 close to natural apatite stoichiometry. Surface profilometry demonstrated that the Ti/HA bilayer coating achieved the highest roughness (12.3 +/- 1.7 & micro;m), which is widely reported in literature to provide a geometrically favorable baseline for potential cellular anchoring. In biotribological tests conducted in simulated body fluid (Lactated Ringer's solution), despite the severe wear loss observed in the brittle single HA layer (77.3 & times; 10-3 mm3 center dot N-1 center dot m-1), the incorporation of the Ti interlayer halved the wear rate in the bilayer system (38.3 & times; 10-3 mm3 center dot N-1 center dot m-1), providing notable mechanical stability. Potentiodynamic polarization measurements confirmed that the Ti/HA bilayer coating acted as a superior electrochemical barrier, exhibiting the lowest corrosion current density (0.50 nA/cm2) and corrosion rate (0.3 & micro;m/year) compared to both single-layer Ti (11.0 & micro;m/year) and HA (45.0 & micro;m/year) coatings. Consequently, the flame-sprayed Ti/HA bilayer coating on SLM-fabricated CoCrWMo implants represents a highly effective surface engineering strategy that simultaneously optimizes corrosion resistance and wear durability for orthopedic applications.