Platinum Nanoparticles/Poly(isoleucine) Modified Glassy Carbon Electrode for the Simultaneous Determination of Hydroquinone and Catechol


KOÇAK S.

Electroanalysis, cilt.33, sa.2, ss.375-382, 2021 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 33 Sayı: 2
  • Basım Tarihi: 2021
  • Doi Numarası: 10.1002/elan.202060232
  • Dergi Adı: Electroanalysis
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, Aerospace Database, Applied Science & Technology Source, BIOSIS, Chemical Abstracts Core, Chimica, Communication Abstracts, Metadex, Civil Engineering Abstracts
  • Sayfa Sayıları: ss.375-382
  • Anahtar Kelimeler: Catechol, hydroquinone, poly(isoleucine), Pt nanoparticles
  • Manisa Celal Bayar Üniversitesi Adresli: Evet

Özet

A new electrochemical sensor based on Poly(Isoleucine) modified glassy carbon electrode decorated with platinum nanoparticles (Pt/Poly(Isoleucine)/GCE) was developed for sensitive individual and simultaneous determination of hydroquinone (HQ) and catechol (CC). Scanning electron microscopy (SEM), Electrochemical impedance spectroscopy (EIS), Cyclic voltammetry (CV) and Differential pulse voltammetry (DPV) were performed in order to characterize the Pt/Poly(Isoleucine)/GCE nanocomposite. For simultaneous determination of HQ and CC, Pt/Poly(Isoleucine)/GCE showed wide linear range between the 0.01–100.0 μM. The detection limits were 0.006 μM for HQ and 0.005 μM for CC. The Pt/Poly(Isoleucine)/GC electrode exhibited good sensitivity and reliability in the simultaneous electroanalysis of two isomers in PBS of pH 7.5. The modified electrode was used to detect the isomers in naturel samples.